Dosimetric characteristics of 6 MV photons from TrueBeam STx medical linear accelerator: Simulation and experimental data

Nuclear Science and Technology, Vol.9, No. 2 (2019), pp. 37-44  
Dosimetric characteristics of 6 MV photons from TrueBeam STx  
medical linear accelerator: simulation and experimental data  
N. D. Ton1, B. D. Linh1,*, Q.T. Pham2  
1 Institute for Nuclear Science and Technology, 179 Hoang Quoc Viet, Cau Giay, Ha Noi  
2 Department of Radiation Oncology and Radiosurgery, 108 Military Central Hospital, Ha Noi  
*Corresponding author, Email: [email protected]  
(Received 17 October 2019, accepted 14 November 2019)  
Abstract: A TrueBeam STx is one of the most technologically advanced linear accelerators for  
radiotherapy and radiosurgery. The Monte Carlo simulation widely used in many applications in  
various fields such as nuclear physics, astrophysics, particle physics, and medicine. The  
Geant4/GATE Monte Carlo toolkit is developed for the simulation in imaging diagnostics, nuclear  
medicine, radiotherapy, and radiation biology to more accurately predict beam radiation dosimetry. In  
this work, we present the simulation results of the dosimetric characteristics of a 6 MV photon beam  
of TrueBeam STx medical LINAC using Monte Carlo Geant4/GATE. The percentage depth dose  
(PDD), central axis depth dose (Profile) have been simulated and compared with those measured in a  
water phantom for field sizes 10×10 cm2 via the gamma-index method. These results will permit to  
check calculation data given by the treatment planning system.  
Keywords: Geant4/GATE, Radiotherapy, TrueBeam STx, Phase Space file, PDD, Profile.  
anatomical sites, where the electron transport  
approximations from analytic or semi-analytic  
dose calculation algorithms are not accurate  
enough. The long computation time becomes  
a reason to hinder MC simulation from  
widespread use in the routine clinical practice  
[2-4]. However, due to industrial secrets, this  
information is sometimes unavailable to the  
I. INTRODUCTION  
The characteristics of the photon beam  
as Percentage Depth Dose and Profile in a  
medical linear accelerator (Linac) are essential  
for building the beam model and guarantee the  
dose distribution for input into Treatment  
Planning System (TPS) [1]. TrueBeam STx is  
one of the latest generation Linac of Varian It  
has different characteristics with the previous  
Linac such as two modes of photon Flattening  
Filter (FF) and Flattening Filter Free (FFF).  
TrueBeam STx can be used for many forms of  
advanced treatment techniques including  
image-guided radiotherapy (IGRT), intensity-  
modulated radiotherapy (IMRT) and RapidArc  
radiotherapy technology.  
general  
medical  
physics  
community.  
TrueBeam STx Linac of Varian Medical  
Systems is an example. Instead, Phase Space  
files (PhS) for each of the beam have been  
made available to the Varian TrueBeam STx  
users [5]. Although the accuracy of these PhS  
among different TrueBeam STx installations  
was proven [6-8], it is advisable to validate  
them by comparing the dose distributions  
produce in the MC simulation with  
experimental measurements performed on a  
TrueBeam STx Linac. Besides, the  
verification of the geometry in this machine is  
also required.  
Monte Carlo (MC) method is a  
popular method to estimate accurate dose  
distributions  
for  
clinical  
beams  
in  
radiotherapy. Especially, the method is  
important for planning heterogeneous  
©2019 Vietnam Atomic Energy Society and Vietnam Atomic Energy Institute  
DOSIMETRIC CHARACTERISTICS OF 6 MV PHOTONS FROM TRUEBEAM STX …  
Geant4/GATE MC code is one of the  
most popular MC open-source applied  
nowadays in medical physics, especially the  
radiotherapy field [9, 10]. Much scientific  
research has been carried out to accurately  
simulate complex geometry in order to find an  
imaging and Cone Beam computed  
tomography (CBCT) capability. One of the key  
features is the availability of two modes of  
photon beams: FF (Flattening Filter) and FFF  
(Flattening Filter Free). FFF beams delivered  
with ―conventional‖ medical linear accelerators  
have the conical flattening filter removed and  
replaced by a thin copper foil (about 1mm  
thick). This foil is the same for all energies.  
Truebeam STx having 6, 8, 10, 15 MV  
standard FF and 6 MV, 10 MV with the new  
FFF beam for photon beams. FFF beams offer  
a maximum high dose rate of 1400 MU/min for  
6 MV FFF and 2400 MU/min for 10 MV FFF,  
respectively [15-17].  
overall reasonable  
agreement  
between  
calculated and experimental doses for Linac  
[11-14]. In Vietnam, this study is the first one  
to assess the capabilities of Geant4/GATE in  
the prediction of characteristics of the  
TrueBeam STx Linac 6 MV photon beam.  
The aim of this work is to evaluate the  
accuracy of Geant4/GATE in simulating  
characteristics of  
6
MV photons from  
TrueBeam STx Linac. In order to establish this  
goal, PDD and Profile were investigated by  
Geant4/GATE with Varian PhS and were  
measured by a Blue Phantom system (The Blue  
Phantom 2, CC013 ionization chamber,  
RAZOR chamber, and electronic equipment).  
Simulations and experimental results were  
compared in specifications of dose difference  
relative to the maximum dose of the  
measurement, we explain the details of this in  
section III. The gamma index was also  
performed using a 2%/2mm standard of dose-  
difference (DD) and distance-to-agreement  
(DTA).  
B. The Phase Space files  
The TrueBeam STx Phase Space files  
for 6 MV FF with the International Atomic  
Energy Agency (IAEA) format [18] have been  
made by Varian using the Geant4 MC toolkit  
[19]. These files were verified as radiation  
sources to allow users to perform accurate  
simulations, which were recorded immediately  
upstream of the movable at a distance of 73.3  
cm from the isocenter. Therefore, users are  
required to code into their MC application only  
the geometrical details of the components of  
the treatment head below the PhS surface,  
including the jaws and MLC. Each PhS file of  
6 MV FF beam contains 109 original histories  
and 5×107 particles. The initial source is an  
electron beam. The energy spectrum of this  
source is non-Gaussian with a peak at 6.13  
MeV and a tail extending up to 6.35 MeV. In  
order to speed up the accelerator head  
simulation, PhS file stored three types of  
particles including photon, electron, positron.  
The details of the header of each file are  
reported [5]. The simulation has been used not  
only photon but also electron and positron in  
PhS. This allows considering electron and  
positron contaminations from the photon beam.  
II. MATERIALS AND METHODS  
A. The Varian TrueBeam STx  
TrueBeam STx with High Definition  
120-MLC (Multi-Leaf Collimator) has at  
central 32 leaf pairs of 2.5 mm and 28 leaf  
pairs of 5 mm leaf thickness to achieve high  
precise target conformation and minimize the  
penumbra effect, which can be used for  
Stereotactic Radiosurgery (SRS), Stereotactic  
Radiation Therapy (SRT) and Stereotactic  
Body Radiation Therapy (SBRT). The  
TrueBeam STx is also equipped with kV/MV  
38  
N. D. TON, B. D. LINH, Q.T. PHAM  
maximum dose (1.5 cm) scans were made for  
C. Experimental measurements  
FF 6 MV with 10×10 cm2 field size. To avoid  
any ripple effect in the measurement, a PDD  
scan was started from the bottom of the tank  
moving toward the surface of the water. Data  
processing and analysis were performed using  
IBA’s OmniPro Accept with application  
setting for a geometric mean smoothing  
function with a value of 3 mm. Appropriate  
stopping power ratio factors were used for  
electron ionization values to PDD values  
conversion. Measured PDD curve was  
compared to Golden Beam Data (GBD) and  
the reproducibility of the measurement data.  
GBD was provided by Varian, which often use  
The Blue Phantom 2 in a large  
48×48×48 cm3 water phantom a three  
dimensions (3D) scanning system was used  
[20]. Fig. 1 was shown actually a photo of Blue  
Phantom 2 under TrueBeam STx head in the  
experiment. The scan in water was acquired at  
the same source-to-surface distance (SSD) =  
100 cm. Beam scanning and collecting data  
were performed in accordance with  
professional guidelines, such as AAPM Task  
Group [21, 22]. The ionization chamber CC13  
and RAZOR chamber [20] were used for beam  
data collection and dosimetric measurement  
following reference [22] recommendation.  
CC13 is a cylindrical ionization chamber with  
a sensitive volume of about 0.13 cm3 and an  
inner diameter of the outer electrode about 6.0  
mm. The effective point offset in the  
measurement of this chamber is 1.8 mm for 6,  
8 MV and 2 mm for 10, 15 MV photon beams.  
The RAZOR is a compact chamber for  
measurements of small fields and of ranges  
with high dose gradients. The RAZOR has  
sensitive volume and the radius cavity are 0.01  
cm3 and 1.0 mm, respectively.  
for  
commissioning  
measurements.  
A
comparison of the PDD with the Gamma Index  
criteria of 2%/2 mm yielded a gamma pass rate  
of 97%. Therefore, the accuracy of this  
measurement can be considered to be within  
2%/2 mm.  
D. Geant4/GATE  
For many years, the Geant4-based  
GATE MC code has been developed as an  
open-source MC program for nuclear medicine  
simulation, with a focus on PET and SPECT  
imaging [23]. This toolkit allows creating a  
simulation on the basis of simple macro-  
command instead of handling tedious C++  
syntaxes of Geant4 code. It helps a quicker  
learning phase for new users and makes a small  
size of the GATE work folder easy to share  
within the community. Details about the GATE  
capabilities and validation are presented  
elsewhere [23-25].  
A new GATE v8.2 was used for this  
simulation [26]. 6 Varian PhS files of smaller  
size (2 GBs) were imported into GATE and  
used for the downstream of the jaws as a  
source in the Linac. These individual files were  
then concatenated to one large PhS file. After  
exiting the PhS plane, the particle passes  
Fig.1. Photograph of Blue Phantom (IBA,  
Germany) scanning water phantom below  
TrueBeam STx head in the experimental setup.  
Measurements include PDD along the  
central axis and crossline profile at the depth of  
39  
DOSIMETRIC CHARACTERISTICS OF 6 MV PHOTONS FROM TRUEBEAM STX …  
through the second collimator as the Y and X  
jaws and MLC. Data for the material and  
geometry of the Linac components were  
obtained from the TrueBeam STx Monte Carlo  
package [19].  
final assessment of the dose distribution  
quality. For the regions of significant  
disagreement, the Gamma Index value is  
greater than unity that will be apparent relative.  
The gamma pass rate was defined as a  
quotient of the passing points and all points.  
For the global Gamma Index passing criteria of  
2%/2 mm, a good agreement, a high  
Geant4 Electromagnetic physics package  
3 (G4EmStandardPhysics_option3) was used  
for precise dose calculations and particle-  
matter interactions or radiation transport in the  
agreement, and  
a
reasonable agreement  
simulation.  
G4EmStandardPhysics_option3  
between the measured and simulated dose  
distribution were observed with over 99%,  
95% and 90% of the points of PDD and cross-  
plane profile, respectively.  
designed for any applications required higher  
accuracy of electrons, hadrons, and ion  
tracking without a magnetic field. The package  
has been presented in Poon and Arce at al. for  
radiotherapy application [27, 28] and  
recommended in Varian documents [19]. The  
range cuts for gamma, electron, and positron  
are fixed to 0.1 mm in a water phantom, 1 mm  
in the world volume, and 10 mm in TrueBeam  
material volume, respectively.  
III. RESULTS AND DISCUSSION  
6 PhS files stored 3×108 photon,  
electron, and positron particles, which have  
been simulated. Approximately 6×109 particle  
histories from 6 PhS files were performed such  
that the statistical uncertainty in the dose for  
the voxels inside the radiation field was less  
than 0.2% at the depth of maximum water  
phantom. The simulation results take into  
A virtual water phantom with 30×30×30  
cm3 volume was installed at an SSD equal to  
100 cm as the measurement. It is used for the  
MC estimation of the absorbed dose  
distribution. Voxel size was set to 3×3×3 mm2  
for a field size of 10×10 cm2.  
account  
the  
electron  
and  
positron  
contamination in the photon beam.  
A. PDD curve  
E. Gamma Index method  
In this paper, the quantity PDD defined  
as the quotient, expressed as a percentage, of  
the absorbed dose at a predefined depth (dx) to  
the absorbed maximum dose at a fixed  
reference depth of d0 = 1.5 cm, along the  
central axis of the beam. Fig. 2 shows the  
comparison between measured and GATE  
estimated PDD with SSD = 100 cm for a  
10×10 cm2 field for FF 6 MV photon beam and  
the Gamma Index distribution. The maximum  
dose was detected at 1.5 cm of depth in both  
measurement and simulation. The statistical  
uncertainty of bins scoring PDD was between  
0.02% to 0.04% and all bins scoring more than  
50% of the maximum absorbed dose was  
0.02% to 0.2%. The distribution of Gamma  
Data analysis was based on comparisons  
between GATE simulations and measurements  
using the Gamma Index method [29], which  
became a ―gold standard‖ method for the  
comparison in dose distribution [13]. This  
method was conducted with a percentage dose  
difference (ΔD) criteria and distance to  
agreement (DTA) of of 1%/1mm and  
2%/2mm. If the Gamma Index value is greater  
than unity, it indicates a position where the  
agreement between the measured and  
simulated dose maps do not meet the  
predefined criteria. Passing criteria were met if  
the gamma index was no larger than 1. An  
important feature of this method is that the  
40  
N. D. TON, B. D. LINH, Q.T. PHAM  
Index is shown in Fig. 2, there is only one  
dose of the measurement as equation 1. This  
discrepancy is never greater than 2%.mm,  
2%.  
point was larger than 1. The evaluation using  
the Gamma Index with 2%/2 mm criteria for  
PDD obtained was greater than 98%. There is a  
good agreement between the computed and  
measured PDD.  
     
 
 
     |  
|        
(1)  
 
 
Where, D1 and D2 are the value of  
simulated and measured, respectively, D0 is the  
maximum dose of the measurement.  
Fig. 3 shows the percentage dose  
difference of the PDD relative to the maximum  
Fig.2: Comparison of measured (black line) and GATE simulation estimated (red circle) PDD of TrueBeam  
STx FF 6 MV photon beam with SSD =100 cm for 10×10 cm2 field. The distribution of gamma index points  
of PDD (blue plus).  
Fig. 3: Percentage dose difference between the simulated and measured PDD relative to the maximum dose  
of the measurement.  
41  
DOSIMETRIC CHARACTERISTICS OF 6 MV PHOTONS FROM TRUEBEAM STX …  
Fig. 4: Comparison of measured (black line) and GATE simulation estimated (red circle) cross-plane profile at 1.5 cm for  
TrueBeam STx FF 6 MV photon beam for 10×10 cm2 field. The distribution of gamma index points of the cross-plane  
profile (blue plus).  
Fig. 5: Percentage dose difference between the simulated and measured cross-plane profile at 1.5 cm of depth  
relative to the maximum dose of the measurement.  
criteria of 1%/1 mm. Although the gamma  
indices in the penumbra region (at ±5 cm) are  
bigger than those in the inside field, there is  
B. Cross-plane Profile  
The comparison of profile at 1.5 cm in  
the water tank between the simulation and  
measure is shown in Fig. 4. The statistical  
uncertainty of the simulation is in the range of  
0.02% to 0.2%. For the Gamma Index criteria  
of 2%/2 mm, the distribution is presented in  
Fig. 4 and the average pass rates for the cross-  
plane profile was ≥ 94%. This agreement  
notably worsens with the more stringent  
still  
a
reasonable agreement between  
computed and measured the cross-plane  
profile at 1.5 cm depth.  
For cross-plane profiles inside the field  
region, the percentage dose difference is less  
than 1.5%. The result of the percentage dose  
difference is shown in Fig. 5. The differences  
42  
N. D. TON, B. D. LINH, Q.T. PHAM  
for clinical radiotherapy treatment planning,‖  
in the penumbra region (at ±5 cm) are bigger  
than those in the inside field. These  
differences probably represent the number of  
simulated particles in which less number of  
simulated particles will be found in the  
penumbra and result in a big statistical  
fluctuations in MC simulation and a big  
difference relative [30, 31].  
Med. Phys. 25, 1773–1829, 1998.  
[2]. N. Reynaert et al., ―Monte Carlo treatment  
planning for photon and electron beams,‖  
Radiation Physics and Chemistry 76, 643–686,  
2007.  
[3]. E. Spezi and G. Lewis, ―An overview of  
Monte Carlo treatment planning for  
radiotherapy,‖ Radiation Protection Dosimetry  
131, 123–129, 2008.  
IV. CONCLUSIONS  
[4]. O. Chibani and C.-M. Ma, ―On the  
discrepancies between Monte Carlo dose  
calculations and measurements for the 18 MV  
Varian photon beam,‖ Med. Phys. 34, 1206–  
1216, 2007.  
The aim of this work is to validate the  
potential application Geant4/GATE software  
for the Varian TrueBeam STx. In this study,  
the characteristics of 6 MV photons of  
TrueBeam STx include PDD and crossline  
profile, which was simulated based on  
Geant4/GATE using Varian PhS file, and  
Varian manufacturer’s information. A PDD  
curve and beam profile for 10×10 cm2 field  
size in a water phantom using Geant4/GATE  
simulation show a good agreement with  
measured dose data for FF 6 MV photon beam  
produced by the Linac. The percentage dose  
difference and Gamma Index method were  
used for comparison. The agreement between  
simulations and experimental data proved that  
Geant4/GATE can be used for accurate Monte  
Carlo dose estimation.  
[5]. M. Constantin et al., ―Modeling the TrueBeam  
linac using a CAD to Geant4 geometry  
implementation: Dose and IAEA-compliant  
Phase Space calculations,‖ Medical Physics  
38, 4018–4024, 2011.  
[6]. C. Glide-Hurst et al., ―Commissioning of the  
Varian TrueBeam linear accelerator: A multi-  
institutional study,‖ Medical Physics 40,  
031719, 2013.  
[7]. G. P. Beyer, ―Commissioning measurements  
for photon beam data on three TrueBeam  
linear accelerators, and comparison with  
Trilogy and Clinac 2100 linear accelerators,‖ J.  
Appl. Clin. Med. Phys. 14, 273–288, 2013.  
ACKNOWLEDGMENT  
[8]. Rodrigues et al., ―A Monte Carlo simulation  
framework for electron beam dose calculations  
using Varian Phase Space files for TrueBeam  
Linacs‖, Medical Physics 42, 2389-2403,  
2015.  
The authors would like to thank  
VINATOM for the support under the grant  
number CS/19/04-02, MSc. Nguyen Ngoc  
Quynh at INST for his kind help in the run of  
Geant4/GATE and all medical physicist at  
Department of Radiation Oncology and  
Radiosurgery, 108 Military Central Hospital.  
[9]. S. Agostinelli et al., ―Geant4—a simulation  
toolkit‖, Nucl. Instr. and Meth. in Phys. Res. A  
506, 250, 2003.  
[10].J. Allison et al., ―Recent developments in  
Geant4‖, Nuclear Instruments and Methods in  
Physics Research A 835, 186, 2016.  
REFERENCES  
[1]. B. Fraass et al., ―American Association of  
Physicists in Medicine Radiation Therapy  
Committee Task Group 53: Quality assurance  
[11].L. Grevillot et al., ―Simulation of a 6 MV  
Elekta precise linac photon beam using  
43  
DOSIMETRIC CHARACTERISTICS OF 6 MV PHOTONS FROM TRUEBEAM STX …  
GATE/GEANT4, Physics in Medicine &  
Biology 56 903–918, 2011.  
[22].R. J. Watts et al., ―Accelerator beam data  
commissioning equipment and procedures:  
Report of the TG-106 of the Therapy Physics  
Committee of the AAPM,‖ pp. 4186–4215,  
2008.  
[12].J. EL Bakkali et al., ―Validation of Monte-  
Carlo Geant4 code for Saturne 43 LINAC‖,  
Int. J. Innov. Appl. Stud. 4, 424–436, 2013.  
[23].S. Jan et al., ―Gate: a simulation toolkit for  
PET and SPECT‖, Physics in Medicine &  
Biology 49(19), 4543-61, 2004.  
[13].S. Didi et al., ―Simulation of the 6 MV Elekta  
synergy platform linac photon beam using  
Geant4 application for tomographic emission‖,  
Medical Physics 40,136–143, 2015.  
[24].S. Jan et al, ―GATE V6: a major enhancement  
of the GATE simulation platform enabling  
modelling of CT and Radiotherapy‖ Physics in  
Medicine & Biology 56(4), 881-901, 2011.  
[14].Wijesooriya et al., ―Verification of the  
TrueBeam Head Shielding Model in Varian  
VirtuaLinac via Out-of-field Doses‖, Medical  
Physics 46(2), 877-884, 2019.  
[25].L. Grevillot et al., ―GATE as a GEANT4-  
based Monte Carlo platform for the evaluation  
of proton pencil beam scanning treatment  
plans‖, Phys. in Med. & Bio. 57(13), 4223-44,  
2012.  
[15].R. Shende et al., ―Commissioning of  
TrueBeam TM Medical Linear Accelerator:  
Quantitative and Qualitative Dosimetric  
Analysis and Comparison of Flattening Filter  
(FF) and Flattening Filter Free (FFF) Beam,‖  
no. February, pp. 51–69, 2016.  
March 2019.  
[16].C. Glide-Hurst et al., ―Commissioning of the  
Varian TrueBeam linear accelerator: a multi-  
institutional study.,‖ Med. Phys., vol. 40, no. 3,  
p. 31719, 2013.  
[27].E. Poon and F. Verhaegen, ―Accuracy of the  
photon and electron physics in Geant4 for  
radiotherapy applications‖, Medical Physics  
32, 1696–711, 2005.  
[17].G. P. Beyer, ―Commissioning measurements  
for photon beam data on three TrueBeam  
linear accelerators, and comparison with  
Trilogy and Clinac 2100 linear accelerators.,‖  
J. Appl. Clin. Med. Phys., vol. 14, no. 1, p.  
4077, 2013.  
[28].P. Arce, JI lagares, ―CPU time optimization  
and precise adjustment of the Geant4 physics  
parameters for a VARIAN 2100 C/D gamma  
radiotherapy linear accelerator simulation  
using GAMOS‖, Physics in Medicine &  
Biology 63(3), 035007, 2018.  
[18].International  
Technical Report IAEA-NDS-0484: ―Phase-  
Space Database for External Beam  
Atomic  
Energy  
Agency  
[29].D. A. Low et al., ―A technique for the  
quantitative evaluation of dose distributions‖,  
Medical Physics 25(5), 656-661, 1998.  
Radiotherapy‖, IAEA Headquarters, Vienna,  
Austria, 2006.  
[30].Graves, Y.J., Jia, X. & Jiang, ―Effect of  
statistical fluctuation in Monte Carlo based  
photon beam dose calculation on gamma index  
evaluation‖, Physics in Medicine and Biology  
58(6): 1839-1854, 2013.  
[19].―TrueBeam Monte Carlo Data Package‖ –  
Confidential document for TrueBeam STx  
users, version 1.1, 2014.  
March 2019.  
–
accessed:  
[31].Saidi, P., Tenreiro, C. & Sadeghi, ―Variance  
Reduction of Monte Carlo Simulation in  
Nuclear Engineering Field‖, Rijeka, Croatia:  
INTECH Open Access Publisher: Theory and  
Applications of Monte Carlo Simulations,  
2013.  
[21].P. R. Almond and P. J. Biggs, ―AAPM‖ TG-51  
protocol for clinical reference dosimetry of  
high-energy photon and electron beams,‖ no.  
September, pp. 1847–1870, 1999.  
44  