Preparation of ¹⁸F-NaF radiopharmaceuticals using home-made automatic synthesis module at Hanoi Irradiation Center

Nuclear Science and Technology, Vol.9, No. 1 (2019), pp. 41-49  
Preparation of 18F-NaF radiopharmaceuticals using home-made  
automatic synthesis module at Hanoi Irradiation Center  
Tran Manh Thang1, Mai Van Vinh1, Dam Thi Tam1, Nguyen Quang Anh1,  
Le Thi Thu Hien1, Pham Minh Duc1, Nguyen Tuan Anh1, Nguyen Van Sy1,  
Nguyen Thanh Hung1, Dang Quang Bao1, Nguyen Xuan Vinh1, Nguyen Thi Kim Dung2  
1Hanoi Irradiation Center. Km12, Road 32, Minh Khai, Tu Liem North, Hanoi, Vietnam  
2Nuclear Medicine Department, 108 Military Central Hospital,  
No. 1, Tran Hung Dao, Hai Ba Trung, Hanoi, Vietmam  
(Received 02 July 2019, accepted 23 September 2019)  
Abstract: The aim of this study is to prepare 18F-NaF radiopharmaceutical using a home-made  
automatic synthesis module consisting of hardware and software which was made by a researcher-  
team of Hanoi Irradiation Center, HIC, Hanoi, Vietnam. Fluorine-18 isotopes produced in cyclotron  
KOTRON13 were transferred to the module and radioactive cation impurities were first removed by  
cation exchange on a carboxymethyl cation exchange (CM) cartridge, and then 18F- ion were trapped  
by a quaternary methyl ammonium anion exchange (QMA) cartridge. Finally, 18F- was eluated from  
the cartridge by isotonic saline water (NaCl 0,9% in water) in the form of 18F-NaF. Time of the  
preparation process was about 13 minutes. Radiochemical yield of the preparation was as high as  
95.5%, in average. The qualities of the product were satisfied the criteria of the United States  
Pharmacopoeia (USP38). PET/CT bone scaner (skeletal scintigraphy) pre-clinical tests using of the  
18F-NaF product showed good quality of imaging for the entire skeletal and distribution of the 18F-NaF  
in the kidney and the bladder agreed with it’s natural distribution.  
Keywords: Sodium [18F]-fluoride radiopharmaceutical, KOTRON13 cyclotron, Synthesis module,  
bone scan.  
reactors. Therefore until 2000, FDA approved  
I. INTRODUCTION  
18F-NaF as a radiopharmaceuticals for bone  
scintigraphy as part of its modernization on  
The  
use  
of  
18F-NaF  
the handling of new drug applications [1]. The  
demand of 18F-NaF radiopharmaceutical is  
expected to increase rapidly in the coming  
years [5]. Nowadays, 99mTc-methylene  
diphosphate (99mTc-MDP) and 18F-NaF have  
been used in bone scintigraphy with SPECT  
(gamma camera) and PET/CT, respectively.  
However, comparing image using 18F-  
NaF/PET-CT technique with those of using  
99mTc-MDP/SPECT showed that the former  
technique has a higher sensitivity in detecting  
osteolytic lesions compared to the latter one.  
The 18F-NaF/PET-CT shows high sensitivity  
in both lytic and blastic lesions. Further, there  
are many advantages of using 18F-NaF in  
radiopharmaceuticals for bone scintigraphy  
dates back to the early 1960s [1]. It was  
introduced as a skeletal PET imaging agent by  
18  
Blau in 1962 [2,3]. The indication of F-NaF  
as a bone scintigraphy agent to define areas of  
altered osteogenic activity was approved by  
the U.S. Food and Drug Administration  
(FDA) in 1972 [2,4]. However, due to the  
unavailability of cyclotron to produce 18F  
preparation in many countries at that time,  
that the bone scintigraphy has been  
conveniently performed with phosphate  
compounds labelled with technetium-99m.  
The 99mTc obtained as a decay product from  
molybdenum-99 [5,6] produced on nuclear  
©2019 Vietnam Atomic Energy Society and Vietnam Atomic Energy Institute  
PREPARATION OF 18F-NAF RADIOPHARMACEUTICALS USING HOME-MADE AUTOMATIC…  
imaging such as sensitivity, specificity, high  
spatial, contrast resolution compared to the  
99mTc use in combination with CT or MRI  
surgery information [7]. In recent years, some  
of nuclear reactors around the world which  
under installation in Kien Giang hospital, two  
in Ho Chi Minh City). Therefore, the  
development and application of 18F-NaF  
radiopharmaceuticals are feasible. In the  
future, it can be replaced a part of 99mTc-MDP  
in the bone cancer diagnosis.  
regularly  
supply  
molybdenum-99  
(technetium-99m generators) are going to  
shutdown [8]. This could be leading to a large  
number of patients would be denied essential  
imaging diagnosis and the nuclear medicine  
community would fall into crisis by the lack  
of 99mTc supply [5]. Moreover, fluorine-18  
isotope that is produced by cyclotron has a  
short half-life (109.83 minutes) so it can not  
be imported from abroad. Up to present  
(2019), Vietnam has seven cyclotrons (three  
in Hanoi City, one in Da Nang hospital, one is  
At present, many companies have  
commercialized  
18F-NaF  
radiopharmaceuticals synthesis module with  
the yield of over 90% such as Eckert &  
Ziegler [9], GE, IBA [1,10], TRASIS [11],  
KIRAMS, etc. Current process of preparation  
of 18F-NaF radiopharmaceuticals worldwide  
is based on the same principle according to  
the flow chart shown below.  
Fig. 1. Flow chart of 18F-NaF radiopharmaceuticals preparation  
In Fig. 1 CM and QMC stand for  
carboxymethyl cation ion exchange and  
quaternary methyl ammonium anion exchange  
radiochemical yields. The 18F-NaF product  
produced was checked for its quality and pre-  
clinical tests.  
In this study, a procedure of 18F-NaF  
radiopharmaceuticals preparation on a home-  
made automatic synthesis module is presented.  
The module was manufactured by a researcher-  
team of Hanoi Irradiation Center (HIC) which  
includes hardware and software to control  
operation of the module (Fig. 2). The process  
of preparation was evaluated for its chemical,  
II. EXPERIMENT  
A. Chemicals, materials and methods  
Material for 18F-NaF module manufacture  
The hardware of the module includes:  
two stepper motor moving up and down, six  
of 2/2 way valves type 0127 ID 120721 from  
42  
TRAN MANG THANG et al.  
Burkert-Germany, six of 3/2 way valves type  
and quick-connect Luer adapters from  
Kinesis-UK, etc. The size of the module is  
400(H) x 300(L) x 285(W) mm (Fig.2).  
Software for operational control of the  
0127 ID 120433 from Burkert-Germany, two  
radioactivity monitoring detectors, Teflon  
tube, PEEK and super flangeless fittings,  
flangeless fittings, flangeless fitting for 1/16”  
OD tubing, bulk head unions, Y connectors,  
module  
was  
written  
on  
Labview  
programming language.  
Fig. 2. A view of the home-made automatic module used for preparation of 18F-NaF  
Chemicals and materials for 18F-NaF  
synthesis  
pH of the preparation was measured  
using indicator strips (Macherey-Nagel);  
half-life of 18F isotope was determined by a  
Dose calibrator (ISOMED2010-Germany);  
Chemical purity of 18F-NaF preparation was  
analyzed by Radio-TLC development  
following by scanning using a Raytest  
scanner from Germany and Radio-HPLC  
Agilent 1260-Raytest; radionuclide purity of  
18F-NaF preparation was identified by using  
18  
Enriched [18O]-water (H2 O) was  
supplied from the Rotem/Israel (18O purity >  
97%). Quaternary methyl ammonium Sep-Pak  
Accell Plus Light (QMA) cartridges and  
carboxymethyl Sep-Pak Accell Plus Short  
(CM) Cartridges were purchased from the  
Waters (USA). Ethyl alcohol (PA grade),  
acetonitrile (PA grade), glass-based silica thin  
layer chromatography (TLC) plates, Millex-  
FG filters were from the Merck/Germany.  
AEF filters 0,22µm pore size are from the  
PALL (UK); sterile water for injection and  
NaCl isotonic saline water (0,9 %) were from  
the Fresenius Kabi Bidiphar/Vietnam.  
Syringes of 20mL capacity were from the BD  
(Belgium). Venting needles were from the  
Millipore (USA) and sterile vacuum vials  
were from Korea.  
multi-channel  
analyzer  
(MCA/Raytest-  
Germany); Residual solvent (ethyl alcohol)  
was determined by Gas chromatography  
(GC/Agilent 7890B); Bacterial Endotoxins  
was tested by Endotoxin PTS-100 (Charles  
River, USA); Sterility was examined in the  
Dalat  
Nuclear  
Research  
Institute’s  
laboratory.  
Quality of 18F-NaF radiopharmaceuticals  
was evaluated at the HIC, Cyclotron Center of  
the 108 Military Central Hospital, and Dalat  
Nuclear Research Institute laboratories.  
Materials and equipment for product’s  
quality control  
43  
PREPARATION OF 18F-NAF RADIOPHARMACEUTICALS USING HOME-MADE AUTOMATIC…  
Ossein scintigraphy was performed with  
the use of 18F-NaF at a PET/CT Light Speed  
device (GE, USA) at Nuclear Medicine  
Department, 108 Military Central Hospital.  
5 mL of enriched 18O-water as a liquid  
target was cỉrculated through a Nb cavity  
covered with Havar foil window. The target  
was bombarded with proton beam of a current  
of 30 µA for 60 minutes. Resulting solution  
containing H18F was transferred into the  
automatic synthesis module (Fig.2, 3). The  
H18F solution first was allowed to pass through  
the CM cation exchange cartridge that was  
prepared previously to purify all cations  
impurities generated from activation reactions  
between protons and the metals made of the  
Havar alloy, then the solution was passed  
through QMA cartridge where 18F- anions were  
trapped on the resin. Finally, 18F was eluted  
from the QMA cartridge with 10 mL of 0,9%  
NaCl isotonic saline solution. The solution of  
18F-NaF was filtered through an AEF filter of  
0.22 µm pore size into a sterile, pyrogen-free  
vial. Progress of the synthetic processes will be  
shown on screen by an animation scheme (Fig.  
3).  
Preparation of QMA, CM cartridges  
before its use  
Sep-park QMA and CM cartridges need  
to be activated before its use. First, both of  
them were preconditioned with 10 mL ethyl  
alcohol then, they were washed by 12 mL  
sterile water for injection. Finally, they were  
dried by blowing a dry air flow.  
Production  
of  
18F-NaF  
radiopharmaceutical  
Fluorine-18  
was  
produced  
by  
bombarment of protons particles accelerated to  
energy 13 MeV on a KOTRON13 cyclotron  
(Korea) located in the HIC. The reaction was  
proceeded as follows.  
188O11p198F01n  
(1)  
Fig. 3. Graphical screen showing progress of 18F-NaF preparation process  
44  
TRAN MANG THANG et al.  
Operation of the module was programed with following menu:  
Section  
Image  
Display  
Operation  
1
2
Value of measured radioactivity will be displayed as graphs  
This menu is for input of necessary data  
Set-up  
3
Control mode  
Process mode  
Start  
This menu is used for automatic control or manual control  
This menu is used for synthesis or cleaning equipment  
Start a process under the conditions which were input  
Stop a process  
4
5
6
Stop  
7
Program running Show current status of the module  
8
Elapsed time  
The Yield  
Display elapsed time since the moment of program started  
9
Display the yield of a synthetic batch  
Display status of process  
10  
Control status  
(2)  
IV. RESULTS AND DISCUSSIONS  
Ten (10) batches of 18F-NaF were  
prepared for evaluating performance of the  
module, and for determining the quality of the  
product and for pre-clinical test with PET/CT  
technique.  
Where Rp is the final product activity measured  
in position of the product storing vial (Fig.3);  
R0 is the initial activity measured in position of  
the QMA cartridge (Fig. 3).  
The yield of 18F-NaF preparation achieved  
in average around 95.5% in 10 experimental  
batches as shown in Fig. 4. The time of the  
synthetic process is about 13 minutes.  
The yield of synthetic process  
The yield of synthetic process was  
estimated by the formula:  
Batch number  
Fig. 4. The yield of 18F-NaF synthetic process determined by batch  
contains no particulate matters. Other  
Quality of the product  
The quality of the product was  
examined. Result of the examination showed  
that 18F-NaF solution was clear, colorless,  
parameters were also satisfied the criteria of  
the United States Pharmacopoeia (USP38)  
applied to 18F-NaF pharmaceuticals as shown  
in Table I [12].  
45  
PREPARATION OF 18F-NAF RADIOPHARMACEUTICALS USING HOME-MADE AUTOMATIC…  
Table I. Quality of 18F-NaF produced at HIC, Hanoi, Vietnam  
Parameter  
Appearance  
Release Criteria (USP38)  
Result  
Clear, colorless, no contains  
particulate matters  
Pass  
pH  
4.5 - 8.0  
> 95%  
7.0  
> 95%  
Pass  
Pass  
Pass  
Pass  
Pass  
Radiochemical purity  
Radionuclidic purity  
Half-life  
> 99,5 % (511±15 keV)  
105-115 min  
< 5 mg/mL  
Residual solvent  
Endotoxin  
< 175 EU/mL  
Sterile  
Sterility  
pH value of the solution was around  
7.0 that met the acceptance criteria of  
between 4.5 to 8.0. Radiochemical purity of  
the product was higher than 95 % of the  
total radioactivity measured as it was  
determined by TLC development. The  
retention factor (Rf) of the product was  
found to be from 0.0 to 0.01 (Fig.5). The  
acceptance criteria for radiochemical purity  
is not less than 95%.  
C/mm *1000  
TLC  
14.0  
12.0  
10.0  
8.0  
6.0  
4.0  
2.0  
0.0  
0
20  
40  
60  
80  
mm  
Fig. 5. The radio-TLC spectrum of 18F-NaF  
Fig.6. The radio-HPLC spectrum of 18F-NaF  
produced at the HIC  
produced at the HIC  
Radiochemical purity and identity of 18F-  
Bacterial Endotoxin was not more than 175  
EU/mL (USP Endotoxin Units per mL) of  
injection that met the acceptance criteria of not  
more than 175 EU/mL. The test for sterility is  
carried out under aseptic conditions. Sterility was  
tested in the fluid thioglycolate medium at 30-  
35oC, the soya-bean casein medium and incubated  
at 20-25oC, respectively. The samples were  
counted 14 days after incubation. The produce was  
sterile that met the acceptance criteria of sterile.  
NaF product were higher than 95 % of the total  
radioactivity measured at retention time around  
4.5 minutes (Fig.6).  
Radionuclidic purity was checked using a  
multi-channel gamma spectrometer and it was  
found that the purity of the product is not less  
than 99.5 % at 511 keV peak (Fig.7) that met the  
acceptance criteria of not less than 99.5%.  
46  
TRAN MANG THANG et al.  
180  
160  
CPS/keV  
Tr1  
140  
120  
100  
80  
60  
40  
20  
0
0
500  
1000  
1500  
keV  
Fig.7. A gamma spectrum of 18F-NaF showing a peak at 511 keV  
Fig.8. A GC chromatogram of solution 18F-NaF produced at the HIC  
showing no peak representing ethyl alcohol  
Fig. 9. PET/CT images of rabbits administered with 18F-NaF preparation produced at the HIC  
Imaging Study and biodistribution  
47  
PREPARATION OF 18F-NAF RADIOPHARMACEUTICALS USING HOME-MADE AUTOMATIC…  
Pre-clinical study is carried out at  
Nuclear Medicine Department, 108 Military  
Central Hospital in Hanoi. Subject of the  
test were rabbits to which were injected  
about 0.3-0.4 mCi (about 0.14 mCi/kg) of  
18F-NaF solution. The uptake time for  
animals was 40 minutes. Afterwards,  
animals were scaned with a low-dose CT,  
full body PET scanner. PET attenuation  
images were corrected by CT. Standardized  
uptake values (SUV) are the maximum  
values at the region of interest and  
calculated in g/mL.  
Institute is acknowledged. The authors would  
like to thank Dr. Jung Young Kim, Korea  
REFERENCE  
[1]. Natalia M. Leonardi, Guillermo A. Casale,  
Jorge Nicolini, Patricia D. Zubata, and Marıa J.  
Salgueiro.  
chromatographic  
alternative for  
Validation  
of  
a
paper  
an  
methodology  
determination of the  
radiochemical purity of Na18F, Nucl. Med.  
Technol., 40: 271-274, 2012.  
as  
Images of 18F-NaF showed that  
metabolism of the radiopharmaceuticals spread  
throughout the entire skeletal of the animal: in  
the skull bone, ribs bones, spine bones, upper  
extremities, two lower limbs, and pelvis.  
[2]. Yuan H., Merrill J. R., Parrott M. and  
Anzellotti A. [18F]NaF Manufacture via the  
ABT  
BG75  
system.  
Imaging  
and  
Biodistribution studies.” University of North  
Carolina at Chapel Hill, NC27599, School of  
Medicine, pp.1-6.  
18  
Distribution of F-NaF product in kidney and  
bladder shows it’s natural distribution.  
[3]. Mark S. Jacobson, Raymond A. Steichen, and  
Patrick J. Peller. PET radiochemistry and  
radiopharmacy",  
Heidelberg, 703: 19-30, 2012.  
Springer-Verlag  
Berlin  
III. CONCLUSIONS  
[4]. Rajeev Kumar, Rajendra G. Sonkawade,  
Madhavi Tripathi, Punit Sharma, Priyanka  
Gupta, Praveen Kumar, Anil K. Pandey,  
Chandrasekhar Bal, Nishikant Avinash Damle,  
and Gurupad Bandopadhayaya. Production of  
the PET bone agent 18F-fluoride ion,  
simultaneously with 18F-FDG by single run of  
the medical cyclotron with minimal radiation  
exposure - A novel technique, Hell J. Nucl.  
Med., 17: 106-110, 2014.  
In this study, it was proven that the Hanoi  
Irradiation Center has successfully manufactured  
automatic synthesis module for 18F-NaF  
radiopharmaceuticals. The duration of the  
synthetic process was about 13 mins.  
Radiochemical yield was as high as 95.5%, in  
average. Quality of the product satisfies the criteria  
of the United States Pharmacopoeia (USP38). The  
PET/CT imaging test on animals showed good  
quality for entire skeletal, kidney and bladder.  
[5]. Brian G. Hockley and Peterj H. Scott. An  
automated method for preparation of [18F]  
sodium fluoride for injection, USP to address  
the technetium-99m isotope shortage. Appl.  
Radiat. Isot., 68: 117-119, 2010.  
ACKNOWLEDGMENTS  
The financial support to conduct this  
research was from Ministry of Science and  
Technology of Vietnam under Project encoded  
ĐTCB.07/16/TTCX. The authors are thankful  
to Vietnam Atomic Energy Institute, and Hanoi  
Irradiation Center for the administrative  
assistances. The co-operation of 108 Military  
Central Hospital and Dalat Nuclear Research  
[6]. Jae Yong Choi, Ji Woong Lee, Kyo Chul Lee,  
Kyeong Min Kim,  
Young  
beom Seo,  
Jung Young Kim, Young Hoon Ryu. Time and  
cost effective production of sodium  
[18F]fluoride using a dedicated automation  
module with disposable cassettes for GMP  
48  
TRAN MANG THANG et al.  
environment. J. Radioanal. Nucl. Chem., 309:  
938-987, 2016.  
[10].Bernard Lambert, Jean-Jacques Cavelier,  
Guillaume Gauron, Christophe Sauvage,  
Cécile Kech, Tim Neal, M. Kiselev,David  
Caron, Anat Shirvan, Ilan Ziv. Routine  
[7]. Frederick D. Grant, Frederic H. Fahey, Alan B.  
Packard, Royal T. Davis, Abass Alavi, and S.  
Ted Treves. Skeletal PET with 18F-Fluoride:  
Applying new technology to an old tracer. The  
J. of Nucl. Med., 49(1): 68-78, 2008.  
automated  
production  
of  
18F-labelled  
radiopharmaceuticals on IBA Synthera®  
Multi-Purpose Platform. Presented at the 13th  
WTTC 2010 in RISOE/Denmark., Abstract  
No.016, 2010.  
[8]. Nuclear and Radiation Studies Board Division  
on Earth and Life Studies. Opportunities and  
approaches for supplying molybdenum-99 and  
associated medical isotopes to Global Markets.  
In Proc of a Sym. Washington, DC. The  
National Academies Press. Chapter 3: Current  
molybdenum-99 supply, pp. 13-17, 2018.  
Langle. Fully automated production of sodium  
[18F]fluoride on All In One and mini All In  
One synthesizers. Appl. Radiat.and Isot., 102:  
87-92, 2015.  
[9]. Modular-Lab PharmTracer, Cassette-based  
routine production of 18F tracers, Eckert &  
Ziegler,  
[12].The United States Phamacopeial Convention,  
“Sodium FluorideF-18 Injection”, USP38, pp.  
3546-3547, 2015.  
(7131-0077/Rev.04/05. 2019). Accessed on 19  
November 2019.  
49  
pdf 9 trang yennguyen 178500 Free
Bạn đang xem tài liệu "Preparation of ¹⁸F-NaF radiopharmaceuticals using home-made automatic synthesis module at Hanoi Irradiation Center", để tải tài liệu gốc về máy hãy click vào nút Download ở trên

File đính kèm:

  • pdfpreparation_of_f_naf_radiopharmaceuticals_using_home_made_au.pdf