Determination of thermal neutron flux distribution at the rotary rack served for elemental concentration analysis using the k₀-INAA method

Nuclear Science and Technology, Vol.10, No. 1 (2020), pp. 11-20  
Determination of thermal neutron flux distribution  
at the rotary rack served for elemental concentration analysis  
using the k0-INAA method  
Nguyen Thi Tho1, Tran Tuan Anh1, Trinh Van Cuong1, Ho Van Doanh1,  
Tran Quoc Duong1, Ho Manh Dung2  
1Dalat Nuclear Research Institute, 01 Nguyen Tu Luc Street., Da Lat, Lam Dong.  
2Center For Nuclear Techniques, 217 Nguyen Trai street, District 1, HCM City.  
(Received 08 January 2020, accepted 24 February 2020)  
Abstract: The accuracy of elements concentration determination using the k0-standardization method  
directly depends on irradiation and measurement parameters including Non-1/E epithermal neutron  
 
flux distribution shape  (  
 
  ), thermal-to-epithermal neutron flux ratio f, efficiency , peak  
⁄
    
 
area… In the case of the irradiation position at the rotary rack of the Dalat Nuclear Research Reactor  
(DNRR), the difference of thermal neutron flux between the bottom (3.541012 n.cm-2.s-1) and the top  
(1.931012 n.cm-2.s-1) of the 15 cm aluminum container is up to 45%. Therefore, it is necessary to  
accurately determine above-mentioned parameters in the sample irradiation position. The present  
paper deals with the determination of the distribution of thermal neutron flux along the sample  
irradiation container by using 0.1% Au–Al wire activation technique. The thermal neutron flux was  
then used to calculate the concentration of elements in the Standard Reference Material 2711a and  
SMELS type III using k0-INAA method at different positions in the container. The obtained results  
with the neutron flux correction were found to be in good agreement with the certified values. In  
conclusion, the proposed technique can be applied for activation analyses without sandwiching flux  
monitors between samples during irradiations.  
Keywords: K0-standardization method, Dalat nuclear research reactor, neutron spectrum parameters.  
The advantages of k0-NAA in the applications  
are a capability of the determination of multi-  
element with high precision and accuracy as  
well as a minimized sample preparation [1].  
I. INTRODUCTION  
Nuclear analytical techniques have been  
developed for decades. It has been used to  
solve  
environmental  
problems,  
legal  
investigations.... Since March 2012, the DNRR  
has been continuously operated about 100130  
hours per month at a nominal power of 500 kW  
for radioisotopes production, activation  
analysis and other researches. The k0-  
standardization method (k0-NAA) has been  
applied and developed at the DNRR over 17  
years. Its main applications include the studies  
in geology, bio-medicine, material, petroleum,  
archaeology and environment among others.  
At the DNRR, there are three irradiated  
channels used for NAA (Fig.1): (1) The fast  
pneumatic transfer system for very short  
irradiation at the channel 13-2 and thermal  
column (Tirr<45 sec); (2) The pneumatic  
transfer system for short and medium  
irradiations at the 7-1 channel (Tirr = 451200  
sec); (3) The rotary rack with 40 irradiated  
holes placed inside the graphite reflector for  
long irradiation  
(Tirr>20  
min). The  
©2020 Vietnam Atomic Energy Society and Vietnam Atomic Energy Institute  
DETERMINATION OF THERMAL NEUTRON FLUX DISTRIBUTION AT THE ROTARY RACK ...  
experiments for determination of thermal flux  
were carried out at the sixth hole of the rotary  
rack of the DNRR.  
in NAA are parameters such as flux variation  
within a sample and irradiation geometry in the  
container. When the sample is irradiated with  
neutrons, the activation rates depend on the  
geometry effect due to the irradiation position  
within the container, the variation and the  
differences within the irradiation site [4].  
In the determination of elemental  
concentration in unknown samples using k0-  
method, samples and flux monitors were  
simultaneously irradiated together. The flux  
monitors were usually positioned at the top,  
middle and bottom of the containers.  
Therefore, we proposed the technique that can  
be applied for k0-method with suitable  
accuracy without sandwiching flux monitors  
between samples during irradiations. An  
experimental determination of the thermal  
neutron flux in the inner sample irradiation  
container at the rotary rack of the DNRR  
using foil activation technique was undertaken  
in this work.  
Fig. 1. Dalat research reactor cross-section.  
The neutron spectrum parameters of the  
DNRR were reported to be very stable which  
permit the use of k0-NAA method [2].  
Instrumental neutron activation analysis with  
research  
reactors  
has  
some  
special  
characteristics which makes it more attractive  
to use for routine analysis. These include  
multi-element capability, reproducibility of  
the results and independence of the chemical  
state of the element [3,4]. Thermal, epithermal  
and fast neutron fluxes determination are  
useful when characterizing the activation site  
in instrumental neutron activation analysis. In  
this perspective NAA using reactor neutrons  
plays a vital role due to its high sensitivity  
and detection limits for many elements in a  
variety of matrices [5], but these could not be  
achieved without proper knowledge of the  
neutron flux [6].  
II. CONTENT  
A. Subjects and methods  
In the absolute method, the thermal  
neutron flux is given as [7]:  
 
 
   
 
   
   
 
 
 
 
 
 
(1)  
(2)  
   
[
]
 
          
 
 
 
 
   
 
      
 
Where Asp is the specific activity, M is  
It is necessary to measure thermal  
neutron flux distribution at various points in  
sample irradiation container and neutron flux  
monitoring is, therefore, required to be carried  
out regularly for any reactor for analytical  
quality control. This is to guarantee continues  
application of neutron activation analysis since  
the main sources of measurement uncertainty  
the atomic mass,  is the isotopic abundance,  
  is the 2200 m.s-1 (n, γ) cross-section, γ is the  
 
absolute gamma-intensity, Np is the number of  
counts under the full-energy peak during the  
counting time tm, w is the sample weight in  
gram, S = l-exp(-tirr) is the saturation factor  
   
 
with tirr being the irradiation time,       
12  
NGUYEN THI THO et al.  
is the decay factor with td being decay time, C  
= [l-exp(-tm) ]/tm is the measurement factor  
correcting for decay during the measurement  
time tm,  is the decay constant, NA is the  
[11]. This is due to the fact that the epithermal  
flux is shown to deviate from the ideal situation  
with a factor . Eq. (4) should, therefore, be  
modified to take care of the flux-shaping factor  
and thus we have a semi-empirical relationship  
given as:  
Avogadro’s number, p is the full energy peak  
detection efficiency, f is the thermal to  
epithermal neutron flux ratio,  is the  
epithermal neutron flux shape factor.  
 
 
[ ]  
      
(5)  
 
    
 
Where  is the characteristic of the  
reactor irradiation position and was shown [11]  
to be positive or negative depending on the  
reactor epithermal spectrum. Eq.(5) was  
proved to be satisfactory for instrumental  
neutron activation analysis [6,11] and it  
enables the correction of the resonance integral  
to the deviating spectrum. Thus to preserve  
accuracy in the k0-method,  should be known  
when calculating the concentration of an  
element in a sample [6,12].  
For ideal situation Q0 = I0/0, I0 -  
resonance integral for an ideal (assumed 1/E)  
epithermal neutron flux distribution.  
For a non-ideal situation, the Q0(I0) need  
to be modified with an -dependent term. The  
conversions from the tabulated Q0(I0) values to  
Q0() (or (I0()) are given by:  
    
    
 
         
      
 
 
 
]     
 
 
  [  
 
(3)  
 
̅̅̅  
 
 
 
 
   
 
         
 
The k0-standardization method was  
Where ECd is the effective Cd cut-off  
introduced in NAA [13]. In terms of the k0-  
energy ( ECd=0.55 eV in standard conditions)  
methodology,  
adopting  
the  
Hødgdahl  
and Er is the effective resonance energy,  
convention [14], the concentration calculations  
are based on the fundamental equation:  
 
 
 
defined by Ryves [8]. The     a term  
(numerically unity) originates from the  
definition of the epithermal neutron flux in a  
1/E1+ distribution [9,10].  
 
 
⁄
 
 
 
 
         
 
 
 
     
 
 
        
(6)  
 
 
 
         
 
 
 
 
⁄
 
 
(
)
     
For an ideal reactor flux, the slowing  
down neutrons after collision with the  
moderator atoms, show an energy distribution  
 e(E) which varies as E-1. This means that the  
epithermal neutron flux integrated over one  
logarithmic energy interval can be represented  
by a constant  e since:  
With k0 defined as:  
 
       
 
 
    
(7)  
 
 
 
       
 
SMELS type III [13] (a multi-element  
synthetic material producing the long-lived  
radionuclides when irradiated with neutrons)  
and SRM-2711a (the certified reference  
material from National Institute for Standard  
and Technology - NIST) were used to calculate  
the concentration of the elements at different  
positions in the container using k0-INAA.  
 
 
[ ]  
      
(4)  
 
 
However, it was found that applying a  
single comparator method (k0-standardization)  
with reactor neutrons, using Eq. (4) is  
unacceptable from the standpoint of accuracy  
13  
DETERMINATION OF THERMAL NEUTRON FLUX DISTRIBUTION AT THE ROTARY RACK ...  
 
   
    
    
In order to evaluate the laboratory  
performance, the u-score test was used in  
which the u-score is calculated according to the  
following equation:  
     
       
(9)  
 
    
B. Experiment  
For determination of the axial flux  
distribution, the gold wire (Al-0.1%Au, =0.6mm)  
was used. The 15 cm long Au wire was placed at  
the center of the aluminum container. After  
irradiation and suitable decay, it was cut in pieces  
of 5 mm with the weight of about 2 - 6 mg. The  
gold foils (Al-0.1%Au, d=0.1mm) and zirconium  
foils (Zr-99.98%) were also placed at some  
positions 1, 2, 6, 7, 12, 14 cm in order to determine  
the thermal neutron flux and , f values.  
 
   
    
    
            
(8)  
 
 
 
√
   
    
    
Where xlab, ulab, xref, and uref are the  
experimental and reference values and  
uncertainties, respectively [2].  
The relative bias between the  
experimental result and the reference value is  
calculated and expressed as a percentage:  
Fig. 2. Typical aluminum sample irradiation container usually used at the rotary rack of the DNRR.  
Table I. The irradiation, decay, counting times for Au wire/foil, Zr monitor, SRM-2711a, SMELS III.  
Counting time  
SRM/Monitor  
Au wire  
Irradiation time (tirr)  
Decay time (td)  
 12d  
(tm)  
1739h  
1h  
1h/10h  
1h  
Au foil  
34 d  
10  15m  
120m  
Zr monitor  
SRM-2711a  
SMELS III  
3d  
1h/10h  
10h  
912d  
215h  
223h  
1018d  
After an appropriate decay time, the  
Au wire was cut into sixteen pieces of 5mm  
and measured on the HPGe coaxial detector  
1.9 keV at 1332.5 keV. The full-peak  
energy efficiency of the detector was  
determined using standard gamma-ray  
sources of 241Am, 133Ba, 109Cd, 137Cs, 60Co,  
57Co and 152Eu.  
(GMX30190), which has  
a
relative  
efficiency of 30% and energy resolution of  
14  
NGUYEN THI THO et al.  
and the top (1.931012 n.cm-2.s-1) of the 15 cm  
C. Results and discussion  
aluminum container is up to 45%. The  
measured results of  and f at the rotary rack  
of the DNRR were found of 0.088 and 39.5,  
respectively. The values are in good agreement  
with the previous measurements [16].  
The thermal neutron flux distribution in  
the sample irradiation container of the DNRR  
was shown in Table II and Fig. 3. The  
difference of thermal neutron flux between the  
bottom (3.541012 n.cm-2.s-1) of the 0.75 cm  
Table II. The axial thermal neutron flux profile in the sample irradiation container.  
Thermal  
neutron flux  
(n.cm-2s-1)  
Axial position in  
the container  
(cm)  
Thermal  
neutron flux  
(n.cm-2s-1)  
Axial position in the  
container  
Error  
Error  
(n.cm-2s-1)  
(n.cm-2s-1)  
(cm)  
0.75  
1.25  
2.25  
3.25  
4.25  
5.25  
6.25  
6.75  
7.25  
8.25  
9.25  
10.25  
12.25  
13.25  
13.75  
14.25  
3.51E+12  
3.43E+12  
3.48E+12  
3.28E+12  
3.16E+12  
3.08E+12  
2.90E+12  
2.86E+12  
2.77E+12  
2.71E+12  
2.68E+12  
2.55E+12  
2.26E+12  
2.13E+12  
2.09E+12  
1.98E+12  
1.43E+11  
1.39E+11  
1.43E+11  
1.32E+11  
1.29E+11  
1.25E+11  
1.18E+11  
1.17E+11  
1.13E+11  
1.15E+11  
1.09E+11  
1.04E+11  
9.25E+10  
8.63E+10  
8.67E+10  
8.16E+10  
2(*)  
7(*)  
3.35E+12  
2.84E+12  
2.06E+12  
3.51E+12  
3.07E+12  
2.27E+12  
1.32E+11  
1.13E+11  
8.31E+10  
1.38E+11  
1.22E+11  
9.11E+10  
14(*)  
1(**)  
6(**)  
12(**)  
(*) Neutron flux in January 2019. (**) Neutron flux in February 2019.  
4.0E+12  
Thermal flux using Au wire  
Thermal flux using Au foils *  
3.5E+12  
Thermal neutron using Au foils **  
3.0E+12  
2.5E+12  
2.0E+12  
1.5E+12  
0
5
10  
15  
Position (cm)  
Fig. 3. Thermal neutron flux distribution from the bottom to top of the container.  
15  
DETERMINATION OF THERMAL NEUTRON FLUX DISTRIBUTION AT THE ROTARY RACK ...  
The thermal neutron flux distribution in  
the sample irradiation container of the DNRR  
also was simulated by MCNP code,  
simulation results were shown in Fig. 4.  
In this simulation (Fig. 5), it has been  
shown that the difference of thermal neutron  
flux within 2.4 cm of the diameter of the  
aluminum container is about 20% and with  
1.5 cm of the effection diameter of the sample  
box is about 10%. This difference won’t affect  
the accuracy of element analysis using k0-  
INAA when irradiation samples are coaxially  
arranged in the.  
1.1  
MCNP  
1
Exp  
0.9  
0.8  
0.7  
0.6  
0.5  
0.4  
The obtained neutron flux was then  
used to calculate concentration of the  
elements in the SRM-2711a and SMELS III  
using k0-INAA method at different positions  
in the container. The xlab/xref ratios, RB values  
and u-scores were used to evaluate the  
precision of data. The thermal neutron flux  
distribution in the container were shown in  
Table II and Fig. 3. The concentration of  
elements Fe, Cr, Co, Sc in SRM-2711a were  
compared with the reference values in which  
the RB values are less than 5% in both cases  
of sandwiching and linear interpolation at  
difference positions in the container, except  
for Cr were about 12% at the position of 14  
cm. The u-score values were within ±1.64 for  
all elements (see Tables III-IV).  
0
5
10  
Positon (cm)  
15  
20  
Fig. 4. Obtained thermal neutron flux distribution  
from the bottom to top of the container by  
experiment and simulation.  
Fig. 4 shows that within the error range  
the obtained results from experiment are in  
good agreement with simulation results.  
The determination of thermal neutron  
flux distribution along the diameter of the  
container by experiment is quite difficult so  
we have simulated this distribution by MCNP  
code. Simulation result was seen in Fig. 5.  
4.0E+12  
3.9E+12  
3.8E+12  
3.7E+12  
3.6E+12  
3.5E+12  
3.4E+12  
3.3E+12  
3.2E+12  
3.1E+12  
-1.8  
-1.2  
-0.6  
0
0.6  
1.2  
1.8  
Position (cm)  
Fig. 5. Simulation of thermal neutron flux distribution along the diameter of the container by MCNP code.  
16  
NGUYEN THI THO et al.  
Table III. The concentration of elements and uncertainties in mg/kg, xlab/xref ratios, RB and u-score values  
for SRM-2711a at position 2cm in the container.  
RB1 RB2  
(%) (%)  
1
2
1
2
Element  
xlab±ulab  
29621±1235 29258±1223 28200±400  
54.2±3.2 53.5±3.2 52.3±2.9  
10.31±0.55 10.19±0.54 9.89±0.18  
8.6±0.4 8.5±0.3 8.5±0.1  
xlab±ulab  
xref±uref  
xlab/xref xlab/xref  
u-score1 u-score2  
Fe  
Cr  
Co  
Sc  
1.05  
1.04  
1.04  
1.01  
1.04 5.04 3.75  
1.02 3.63 2.29  
1.03 4.25 3.03  
1.00 1.18 0.00  
1.09  
0.44  
0.73  
0.24  
0.82  
0.29  
0.52  
-0.05  
1 using sandwiching flux monitor  
2 using linear interpolation  
Table IV. The concentration of elements and uncertainties in mg/kg, xlab/xref ratios, RB and u-score values  
for SRM-2711a at position 7 cm in the container.  
RB1 RB2  
(%) (%)  
2.08 2.32 0.46  
u-  
u-  
1
2
1
2
Element  
xlab±ulab  
xlab±ulab  
xref±uref xlab/xref xlab/xref  
1.02  
1.00 -0.18 -4.06 0.47  
score1 score2  
Fe  
Cr  
Co  
Sc  
28786±1197 28854±1180 28200±400 1.02  
0.53  
0.5  
54.1±2.6  
10.25±0.47 9.85±0.44  
8.6±0.4 8.3±0.3  
52.0±2.5  
54.2±2.6  
9.89±0.18  
8.6±0.4  
1.00  
1.04  
1.00  
1.00  
0.97  
3.64 -0.40 0.72  
0.00 -3.49 0.32  
0.78  
0.38  
Table V. The concentration of elements and uncertainties in mg/kg, xlab/xref ratios, RB and u-score values for  
SRM-2711a at position 14 cm in the container.  
RB1 RB2  
u-  
2
1
2
1
Element  
xlab±ulab  
xlab±ulab  
xref±uref xlab/xrefxlab/xref  
u-score2  
score1  
(%) (%)  
28726±1234 28985±1223 28200±400 1.02 1.03 1.87 2.78  
58.2±3.2 58.7±3.2 52.3±2.9  
10.16±0.51 10.25±0.51 9.89±0.18 1.03 1.04 2.73 3.64  
8.8±04 8.9±0.4 8.5±0.1 1.04 1.05 3.53 4.71  
Fe  
Cr  
Co  
Sc  
0.41  
0.61  
1.49  
0.67  
0.95  
1.11 1.12 11.28 12.24 1.37  
0.49  
0.73  
Table VI. The concentrations of elements and uncertainties in mg/kg, xlab/xref ratios, and u-scores  
for SMELS III at position 1 cm in the container.  
RB1 RB2  
(%) (%)  
u-  
u-  
1
2
1
2
Element xlab±ulab  
xlab±ulab  
xref±uref  
xlab/xref xlab/xref  
score1 score2  
Fe  
Co  
Cr  
Sc  
Cs  
In  
8672±358 8655±357  
8200±190  
1.06  
1.05  
1.04  
1.06  
1.05  
1.04  
1.06  
1.08  
1.10  
1.07  
1.10  
1.09  
1.11  
1.07  
1.09  
1.07  
5.76 5.55 1.16 1.12  
4.94 4.73 1.09 1.05  
4.04 3.81 0.76 0.72  
6.14 6.14 1.16 1.12  
8.56 8.32 1.80 1.76  
10.28 10.17 1.69 1.66  
7.03 6.64 1.37 1.33  
10.69 9.92 1.61 1.57  
9.31 9.09 1.79 1.75  
10.69 10.69 1.88 1.84  
7.30 7.30 1.40 1.36  
9.18 8.70 1.76 1.72  
6.96 6.80 1.49 1.44  
25.50±1.04 25.45±1.04 24.3±0.33  
90.2±3.79 90.0±3.8 86.7±2.6  
1.21±0.05 1.21±0.01 1.140±0.031 1.06  
22.58±0.92 22.53±0.92 20.80±0.34  
1.09  
1.10  
1.07  
1.11  
1.09  
1.11  
1.07  
1.09  
1.07  
510±21  
54.8±2.2  
145±6  
509±21  
54.6±2.2  
144±6  
462±19  
51.2±1.3  
131±6  
Sb  
Se  
Sr  
8909±375 8891±374  
8150±200  
26.2±0.9  
23.3±0.7  
20.7±0.5  
618±11  
Th  
Tm  
Yb  
Zn  
29±1  
25±1  
29±1  
25±1  
22.6±0.9  
661±27  
22.5±0.9  
660±27  
17  
DETERMINATION OF THERMAL NEUTRON FLUX DISTRIBUTION AT THE ROTARY RACK ...  
Table VII. The concentrations of elements and uncertainties in mg/kg, xlab/xref ratios, and u-scores for  
SMELS III at position 6 cm in the container.  
RB1 RB2  
u-  
u-  
1
2
1
2
Element xlab±ulab  
xlab±ulab  
xref±uref  
xlab/xref xlab/xref  
score1  
score2  
(%) (%)  
1.57 6.06  
0.41 4.94  
4.96 9.57  
0.88 5.26  
0.96 5.77  
5.63 10.39  
1.56 7.42  
5.34 9.92  
3.73 8.33  
7.25 12.21  
9.87 9.68  
9.18 14.01  
0.81 5.18  
Fe  
Co  
Cr  
Sc  
Cs  
In  
8329±485 8697±505  
8200±190  
24.3±0.33  
86.7±2.6  
1.02  
0.96  
1.05  
1.01  
1.01  
1.06  
1.02  
1.05  
1.04  
1.07  
1.10  
1.09  
1.01  
1.06  
1.05  
1.10  
1.05  
1.06  
1.10  
1.07  
1.10  
1.08  
1.12  
1.10  
1.14  
1.05  
0.25  
0.07  
0.82  
0.20  
0.19  
0.91  
0.42  
0.77  
0.64  
1.29  
1.62  
1.53  
0.14  
0.92  
0.90  
1.49  
0.95  
1.02  
1.62  
1.17  
1.42  
1.37  
2.06  
1.77  
2.24  
0.91  
24.4±1.2  
91±5  
25.5±1.3  
95±5  
1.15±0.06 1.20±0.06 1.140±0.031  
21±1  
488±22  
52±3  
22±1  
510±23  
55±3  
20.80±0.34  
462±19  
Sb  
Se  
Sr  
51.2±1.3  
131±6  
138±7  
144±7  
8454±433 8829±453  
28.1±1.2 29.4±1.3  
25.6±1.3 22.7±1.02  
8150±200  
26.2±0.9  
23.3±0.7  
20.7±0.5  
618±11  
Th  
Tm  
Yb  
Zn  
22.6±1.2  
623±32  
23.6±1.2  
650±34  
Tables VI and VII show elements  
concentration and uncertainties in mg/kg for  
SMELS III sample along with xlab/xref ratios,  
RB values and u-scores at different placed  
sample positions in the container with  
sandwiching flux monitors and with linear  
interpolation of thermal neutron flux. These  
results were compared with the obtained results  
by the other authors [18] in which the RB  
values for all elements were within ±5%.  
Generally, most u-score values were within  
±1.96 except for Th, Yb (the second method  
was marked 2, at position 6 cm in the  
container) were bigger than this value (Table  
VII). If we increase the limiting value for the  
u-score to 2.58 for a level of probability at  
99%, all our analytical results will pass. So we  
see that determination of concentration of  
elements using k0-INAA method with or  
without sandwiching flux monitors are  
acceptable. It means that we can get the precise  
of thermal neutron flux at each position of the  
sample from the linear interpolation instead of  
sandwiching the monitor as usual.  
III. CONCLUSIONS  
The axial thermal neutron flux  
distribution in the sample irradiation container  
at the rotary rack of the DNRR was determined  
by gold wire activation analysis. The  
difference of thermal neutron flux within 0.5  
cm of the aluminum container length is about  
2%, 1 cm is about 4%, 2 cm is about 10% and  
between the bottom and the top of the  
aluminum container is up to 45%. The  
accuracy of element analysis using k0-INAA  
method strongly depends on the thermal  
neutron flux. Hence, the proposed technique  
can be applied for the determination of  
concentration of elements without sandwiching  
flux monitors.  
18  
NGUYEN THI THO et al.  
[6]. Ahmed, Y.A., Ewa, I.O.B., Umar, I.M.  
ACKNOWLEDGMENT  
“Effective resonance energy and nonideality of  
epithermal neutron flux distribution in neutron  
activation analysis”, Nigerian Journal of  
Physics, 14(1), 82-85, 2002.  
The authors are thankful to the  
Ministry of Science and Technology of  
Vietnamese government for financial  
support through the ministerial-level project  
DTCB-01/18-VNCHN.  
[7]. H.M Dung. “Study for development of k-zero  
Neutron Activation Analysis for multi-element  
characterization”, PhD thesis, the Natural  
Science University, Hochiminh city, 2003.  
REFERENCE  
[8]. T.B. Ryves. “A new thermal neutron flux  
convention”, Metrologia, 5(4), 119-124, 1969.  
[1]. Freitas MC, Révay Z, Szentmiklosi L, Dionisio  
I, Dung HM, “Different methodologies in  
neutron activation to approach the full analysis  
of environmental and nutritional samples”,  
[9]. F. De Corte, S. Van Lierde. “Determination  
and evaluation of fission k0-factors for  
correction of 235U(n,f) interference in K0-  
NAA”, Journal of Radioanalytical and Nuclear  
Chemistry, 248(1), 97-101, 2001.  
Journal  
Radioanalytical  
and  
Nuclear  
Chemistry, 278(2), 381–386, 2008.  
[2]. C.D. Vu, T.Q. Thien, H.V. Doanh, P.D. Quyet,  
T.T.T. Anh and N.N. Dien, “Characterization  
of neutron spectrum parameters at irradiation  
channels for neutron activation analysis after  
full conversion of the Dalat nuclear research  
reactor to low enriched uranium fuel", Nuclear  
Science and Technology, 4(1), 70-75, 2014.  
[10].F. De Corte, F. Bellemans, P. De Neve, A.  
Simonits. “The use of a modified Wescott  
formalism in the K0-standadization of NAA”,  
Journal of Radioanalytical and Nuclear  
Chemistry, 179(1), 93-103, 1994.  
[11].De Corte, F., Moens, L., Simonits, A.,  
Hammami, K.S., De Wispelaere, A., Hoste,  
J.. “The effect of epithermal neutron flux  
distribution on the accuracy of absolute and  
comparator standardization methods in (n, )  
[3]. Jonah, S.A., Umar, I.M., Oladipo, M.O.A.,  
Balogun,  
G.I.,  
Adeyemu,  
D.J..  
“Standardization of NIRR-1 irradiation and  
counting facilities for instrumental neutron  
activation analysis”, Applied Radiation and  
Isotopes, 64(7), 818-822, 2006.  
activation  
analysis”,  
Journal  
of  
Radioanalytical and Nuclaer Chemistry,  
72(1-2), 275-286, 1982.  
[4]. Ahmed, Y.A., Landsberger, S., O’Kelly,  
D.J., Braisted, J., Gabdo, H., Ewa, I.O.B.,  
Funtua, I.I., Umar, I.M.. “Compton  
suppression method and epithermal NAA  
in the determination of nutrients and heavy  
metals in Nigerian food and beverages”,  
Applied Radiation and Isotopes, 68(10),  
1909-1914, 2010.  
[12].De Corte, F., Simonits, A., De Wispelaere, A.,  
Hoste, J. “Compilation of k-Factors and  
Related Nuclear Data for 94 Radionuclide of  
Interest in NAA INW/KFKI Interim Report”,  
INW/KFKI Interim, 1986.  
[13].A. Simonits, F. De Corte and J. Hoste. “Simple  
comparator Methods in Reactor Neutron  
Activation  
Analysis”,  
Journal  
of  
Radioanalytical and Nuclear Chemistry, 24(1),  
31-46, 1975.  
[5]. Acharya, R.N., Nair, A.G.C., Reddy, A.V.R.,  
Manohar, S.B.. “Validation of a neutron  
activation analysis method using k0  
–
[14].O. T. Hødgdahl. “Neutron absorption in pile  
neutron activation analysis”, Report MMPP-  
226-1, Michiga University, Dec. 1962.  
standardization”, Applied Radiation and  
Isotops, 57(3), 391-398, 2002.  
19  
DETERMINATION OF THERMAL NEUTRON FLUX DISTRIBUTION AT THE ROTARY RACK ...  
[15].Vermaercke P, Robouch P, Eguskiza M, De  
Corte F, Kennedy G, Smodis B, Jacimovic R,  
Yonezawa C, Matsue H, Lin X, Blaauw M,  
Kucera J, “Characterisation of synthetic multi-  
element standards (SMELS) used for the  
validation of k0 –NAA”, Nucl Instrum Meth A  
564(2), 675–682, 2006.  
[17].F. De Corte, A. Simonits. “Recommended  
nuclear data for use in the k0  
standardization of neutron activation  
analysis", Atomic Data and Nuclear Data  
Tables 85(1), 47-67, 2003.  
[18].Wasim M, Zaidi JH, Arif M, "Use of synthetic  
multi-element standards (SMELS) in the  
quality control and method validation of k0-  
neutron activation analysis”, Radiochim Acta,  
98(3), 183-186, 2010.  
[16].H.M. Dung, T.Q. Thien, H.V. Doanh, C.D. Vu,  
N.T. Sy. “Quality evaluation of the k0-  
standardized neutron activation analysis at the  
Dalat research reactor", Journal of Radioanalitical  
and Nuclear Chemistry, 309(1), 235-241, 2016.  
20  