A new rapid neutron activation analysis system at Dalat nuclear research reactor

Nuclear Science and Technology, Vol. 4, No. 1 (2014), pp. 84-91  
A new rapid neutron activation analysis system  
at Dalat nuclear research reactor  
H.V. Doanh*, C.D. Vu, T.Q. Thien, P.N. Son, N.T. Sy, N. Giang and N.N. Dien  
Nuclear Research Institute, 01 Nguyen Tu Luc Street, Dalat, Vietnam  
(Received 5 March 2014, accepted 12 March 2014)  
Abstract: An auto-pneumatic transfer system has been installed at the Dalat research reactor for rapid  
instrument neutron activation analysis based on very short-lived nuclides. This system can be used to  
perform short irradiations in seconds either in the vertical channel 13-2 or in the horizontal thermal  
column of the reactor. The transferring time of sample from irradiation to measurement position is  
approximately 3.2 seconds. A loss-free counting system using HPGE detector has been also setup in  
compacting with the pneumatic transfer system for measurement of sample’s activity, automatically  
starting for data acquisition at irradiated sample’s arrival. This new facility was tested and shown to  
have high potential for the determination of short-lived nuclides with half-lives from 10 100  
seconds. This work presents the results of timing parameter measurements, characterization of  
irradiation facilities, and application of this system to determining Selenium concentration in several  
biological reference materials.  
Keywords: Auto-pneumatic transfer system, neutron activation analysis, short-lived nuclides.  
I. INTRODUCTION  
In the recent years, through the IAEA  
TC Project RER/4/028, a new automatic PTS  
Instrumental neutron activation analysis  
for rapid neutron activation analysis based on  
(INAA) has been developed and applied at the  
short-lived nuclides has been developed. This  
500 kW Dalat research reactor (DNRR) since  
facility consists of three main parts introduced  
1984. Until now, it is capable of analyzing  
in reference [1]. The first part, consisting of  
more than 40 elements based on radionuclides  
two aluminum irradiation tubes, which are  
with short, medium and long-lived time. For  
inserted into the vertical channel No.13-2 and  
short-lived nuclides with half-lives from 2  
the horizontal thermal column (TC) of the  
minutes to 2.6 hours, samples are often  
reactor. The second part is a digital signal  
irradiated at the neutron channel No.7-1 of  
processing spectrometer connected to a 40%  
Dalat research reactor through a semi-auto  
relative efficiency HPGe detector coupled with  
pneumatic transfer system (PTS) with valid  
a transistor reset preamplifier. The third part is  
irradiation time from 45 seconds to 20  
composed of pneumatic chambers, loading and  
minutes. Measurements are often performed  
using a gamma spectrometer coupled with a  
HPGe (GMX-30190), but with manual  
manipulation between loading and counting  
procedures. Therefore, the shortest-lived  
sliding devices in Cabin-1 which facilitates the  
fully automatic irradiation-counting  
procedures. It has also a sample automatic  
loader for the sequential routing of the samples  
in multi-samples operation mode: when the  
measurement for one sample is finished, the  
next sample is loaded and sent to the  
irradiation and then counting positions. In this  
nuclides that could be detected are 28Al (T1/2  
=
2.24 min), 52V (T1/2 = 3.75 min), and 51Ti (T1/2  
= 5.76 min).  
©2014 Vietnam Atomic Energy Society and Vietnam Atomic Energy Institute  
HO VAN DOANH et al.  
system, there is also has optical sensors for  
77mSe, 179mHf, 46mSc, and 110Ag can be used for  
INAA at Dalat reactor, which the former  
system can not detect.  
controlling the transport of the capsule as a  
sample carrier, and for accurately measuring  
the capsule flight time from irradiation position  
to detector. The installation diagram is shown  
in Fig. 1.  
The main purpose of this work is to test  
the system for both mechanical and analytical  
reliability. A systematic study has been carried  
out including measurements for timing  
parameters of the system and neutron flux at  
irradiation positions, and the application of this  
system to determining of Selenium in a number  
of biological reference materials for validation  
purpose.  
This PTS system can be used to perform  
short irradiations in seconds. The return time of  
sample from irradiation position to counting  
position is about 3.2 s. Timing information for  
both irradiation and counting will be instantly  
delivered to the activation analysis workstation  
computer. The digital gamma spectrometer is  
selected and tuned for accurate measurement at  
high and varying counting rates, using loss-free  
counting technology. Accordingly, shorter-  
lived nuclides (half-life < 1 min) such as 20F,  
Fig. 1. Diagram of the auto-pneumatic transfer system installed at DNRR.  
85  
A NEW RAPID NEUTRON ACTIVATION ANALYSIS SYSTEM AT …  
II. EXPERIMENT  
A. Timing measurements  
recorded at the computer via a signal from the  
fast solenoid providing the return gas (START  
of return time) and from the optical sensor at  
the detector station recording the arrival of  
capsule (END of return time).  
The accuracy of irradiation time of an  
irradiation facility should be checked and  
calibrated as a type of analytical qualify  
control [2]. In this experiment, the absolute  
irradiation time, the sample transferring time  
from entrance to irradiation position of  
aluminum tube (Tin) and the reverse movement  
(Tout) were determined. The experiments were  
done outside the reactor before installation of  
irradiation facilities inside the reactor. The  
arrangement of timing measurements is shown  
in Fig. 2. The setup includes two NaI(Tl)  
detectors placed at the top and bottom sides of  
the irradiation tube. The detectors detect  
gamma radiation pulses from a 131I source  
inserted in a capsule while moving inside. The  
counters are set for running in Multi-Channel  
Scaler (MCS) mode; MCS mode records the  
counting rate of events as a function of time.  
B. Neutron spectrum parameters of irradiation  
position  
The neutron spectrum parameters  
including thermal neutron flux th, fast neutron  
flux f, the thermal flux to epithermal neutron  
flux (epi) ratio were measured at sample  
irradiation positions in channel No.13-2 and  
thermal column using Au, Zr, and Ni monitors.  
Monitors were inserted into a high purity  
polyethylene vial and loaded into rabbit  
(capsule) for irradiation. The activity  
measurements were carried out by a calibrated  
gamma-ray spectrometer combined with HPGe  
detector (GMX-30190). The measured  
spectrum were analyzed by using the k0-IAEA  
program. The irradiation, decay and counting  
times for each monitor are shown in Table I.  
Typically monitors with masses of 4 mg for  
Al-0.1%Au foil (IRMM-530R), 30 mg for pure  
Ni (wire), 10 mg for Zr (foil) were irradiated  
for 10 min at 13-2 channel (2 h at thermal  
column), and the decay time is 1 day for 97Zr  
and 3 days for 198Au, 95Zr and 58Co.  
The return time from the irradiation to  
the measurement positions was determined by  
a series of irradiation (50 replicates) for a total  
weight (capsule, vial and sample) of about 4.4  
gram and air pressure of 3.1 bars over a  
distance of 40 m for 13-2 channel and 36 m for  
thermal column. The return times were  
Fig 2. Arrangement for experiment of timing  
86  
HO VAN DOANH et al.  
Table I. The irradiation, decay and counting times for the monitors.  
Time/position  
irradiation (monitor,  
mass)  
Decay  
time  
Counting time  
(combination)  
Measured  
radionuclides  
(T1/2, -rays in keV)  
10 min/ 13-2 channel  
2 hours/ thermal column  
(Al-0.1% Au, ~ 4 mg)  
(99.8% Zr, ~ 10 mg)  
(99.98% Ni, ~ 30 mg)  
~ 1 d  
~ 3 d  
97mNb (60 s, 743.4)*;  
97Nb (16.7 h, 657.9)  
198Au (2.7 d, 411.8);  
95Zr (64 d, 765.8);  
58Co (70.8 d, 810.8)  
1 2 h  
0.5 3 h (5 h)  
* Nuclide 97mNb is decayed from nuclide 97Zr with half-life of 16.7h.  
C. Determination of Selenium  
were irradiated for 25 s, allowed 20 s delay  
time to eliminate interference of 116mIn with a  
half-life of 2.18 s [3, 4]) and counted for 25 s  
at a distance of 10 cm from the detector  
(GMX40-76-PL). The concentrations of  
Selenium were determined by both k-zero and  
relative methods.  
A variety of reference materials (Tuna  
Fish IAEA-436, Oyster tissue NIST 1566b,  
Bovine Liver NIST 1577, Bovine Liver NIST  
1577b) were selected to assess reliability of  
this system on the short-time activation  
application. All of the samples were irradiated  
at a neutron flux of 4.21012 n.cm-2.s-1 in the  
13-2 channel and counted on the calibrated  
HPGe gamma-ray spectrometer (GMX40-76-  
PL).  
III. RESULTS AND DISSCUSION  
A. Timing measurements  
The results for average transferring time  
of sample from the top to bottom of the  
aluminum irradiation tube (Tin) is (0.628   
0.021) s for the channel No.13-2 irradiation  
tube (a length of 6 m) and Tout is (0.323   
0.030) s (averaged for 90 runs over the three  
days). For thermal column irradiation tube (a  
length of 2.8 m), Tin is (0.248 0.019) s and  
Tout is (0.146 0.004) s, as shown in Table II.  
The result obtained for measuring the return  
time from the irradiation position to the  
measurement position was found to be (3.165 ±  
0.002) s for channel No.13-2. That for thermal  
column was (3.025 0.013) s. It should be  
noted that this timing parameters are included  
in the time required to start the detector after  
receiving the start signal.  
In order to evaluate the limit of detection  
of Se in biological samples, two 200mg  
replicates of each material (IAEA 436 and  
NIST 1566b) were weighed and packed in high  
purity polyethylene bags. The samples were  
irradiated for 5, 10, 15, 20, 25, 30, 35 and 40 s.  
After a delay of 3.2 s (including both  
transferring time of sample from irradiation  
position to detector and the time required to  
start the detector). Each sample were counted  
for 20 s at a distance of 10 cm from detector.  
To test accuracy for the analysis of the  
Se concentration in biological reference  
materials, four 200 mg replicates of each  
material (IAEA 436, NIST 1566b, NIST 1577  
and NIST 1577b) were weighed. The samples  
87  
A NEW RAPID NEUTRON ACTIVATION ANALYSIS SYSTEM AT …  
Table II. The result of time measurements.  
Irradiation  
position  
The transferring time throughout  
aluminum irradiation tube (second)  
The return time from irradiation  
position to detector position (second)  
Tin  
Tout  
This word  
Manufacturer*  
13-2 channel  
3.165 ± 0.002  
0.628 0.021  
0.323 0.030  
3.301 0.013  
Thermal column  
0.248 0.019  
0.146 0.004  
3.025 0.013  
3.261 0.022  
* Sample weight: 8 g for thermal column tube and 6 g for 13-2 channel tube,  
operation air pressure: 3.1 bars, distance: 30 meters.  
There are significant differences  
between this work and that of the manufacturer  
in capsule sample weight and distance from  
irradiation position to measurement position.  
Hence, there are differences (7%) in the  
result of the return time from irradiation  
position to detector position. However, it is not  
a problem for analytical measurements.  
channel No.13-2 and 4.91% for thermal  
column. The relative error is less than 1% at  
irradiation time of 5 s for channel No.13-2, and  
10 s for thermal column. The large error for the  
first second is due to delay of the system in  
starting the irradiation timer and in ejecting the  
capsule once the “end of irradiation” signal has  
been received.  
This timing delay problem can be  
adjusted through the control unit and the  
software package for managing optimal  
operation and the analytical procedures.  
However, it is not a problem for INAA because  
the time parameters remain unchanged for all  
samples, standards, and control material  
Results for absolute irradiation time at  
channel No.13-2 and thermal column were  
determined by a series of irradiations ranging  
from 1 to 30 s (3 replicates), as shown in Fig. 3  
and Fig. 4. The relative error of irradiation time  
in the first second is 16.02% for channel  
No.13-2 and 26.43% for thermal column, and  
those for irradiation time of 2 s is 1.5% for  
18  
16  
14  
12  
10  
8
28  
24  
20  
16  
12  
8
6
4
4
2
0
0
0
2
4
6
8
10 12 14 16 18 20 22 24 26 28 30  
Irradiation time, second  
0
2
4
6
8
10 12 14 16 18 20 22 24 26 28 30  
Irradiation time, second  
Fig. 3. The relative error of irradiation time  
for 13-2 channel.  
Fig. 4. The relative error of irradiation time for  
thermal column.  
88  
HO VAN DOANH et al.  
B.Neutronspectrumparametersofirradiationpositions  
energy [5], measured using the 58Ni(n,p)58Co  
nuclear reaction. The thermal neutron flux at  
the irradiation position in the thermal column  
is 1.25E+11 n.cm2.s-1, associated with much  
lower fast and epithermal neutron flux.  
Hence, thermal column is a useful irradiation  
channel for eliminating interference reactions  
induced by fast neutron, in which sample is  
irradiated in an extremely well thermalized  
neutron field [6].  
The results of the determination of  
neutron spectra parameters are shown in Table  
III. This table includes data obtained for the  
thermal, fast neutron flux, the ratio of thermal  
to epithermal neutron flux (th/epi). The  
thermal neutron flux at the irradiation position  
in the channel No. 13-2 is 4.2E+12 n.cm2.s-1,  
and associated with 0.5 times of epithermal.  
The integral fast neutron flux is 6.61E+12  
n.cm-2.s-1 for all neutrons above 2.9MeV in  
Table III. The results of neutron spectra parameters at irradiation positions in the channel No.13-2  
and thermal column of DNRR.  
th (n/cm2/s)  
th /epi  
F (n/cm2/s)  
Irradiation position  
(4.2 0.1) x 1012  
(6.6 0.9) x 1012  
10.7 2.4  
13-2 channel  
(1.24 0.03) x 1011  
(8.4 0.5) x 108  
195 4  
Thermal column  
C. Determination of Selenium  
sensitivities for Se rapid determination in a  
variety of biological matrices.  
Finally, measurements of detection  
limits of Se in IAEA 436 and NIST 1566b  
samples were performed. The results for these  
measurements are presented in Fig 4. The  
obtained results confirm that in irradiation  
from 15 s to 25 s at irradiation position of the  
channel No.13-2 coupled with counting for  
roughly 20 s at 10 cm distance from detector,  
the detection limits for Se is within the range  
0.5 0.7 ppm, depending on the sample  
composition. It provides adequate analytical  
The accuracy for determination of  
Selenium using the short-lived nuclide 77mSe  
was evaluated by analyzing a number of  
certified reference materials with different  
levels of Se (IAEA 436, NIST 1566b, NIST  
1577 and NIST 1577b). The agreement  
between measured and certified values was  
generally very good with u-score < 1.64, as  
shown in Table IV.  
1.8  
1.6  
1.4  
1.2  
1.0  
0.8  
0.6  
0.4  
0.2  
NIST 1566b  
IAEA 436  
0
5
10  
15  
20  
25  
30  
35  
40  
45  
Irradiation time, second  
Fig. 4. The detection limits of Se in IEAE 436 and 1566b.  
89  
A NEW RAPID NEUTRON ACTIVATION ANALYSIS SYSTEM AT  
For the determination of the Selenium  
lived nuclide, not only completion times are a  
distinct advantage but analytical sensitivities  
are also improved. The data for procedures are  
listed in Table V.  
by the instrumental neutron activation analysis,  
the long-lived nuclide 75Se or the short-lived  
nuclide 77mSe can be used [7]. With the short-  
Table IV. The results of concentration analysis for Se in biological reference materials.  
Certificated  
value  
(in ppm)  
k-zero method  
The relative method  
This work  
Reference  
material  
This work  
(in ppm)  
u-score  
0.12  
u-score  
(in ppm)  
IAEA 463  
NIST 1566b  
NIST 1577  
NIST 1577b  
0.66  
0.27  
0.29  
0.39  
4.63 0.48  
2.06 0.15  
1.10 0.10  
0.73 0.06  
4.55 0.50  
4.19 0.46  
0.71  
2.48 0.57  
1.24 0.31  
0.70 0.11  
2.18 0.42  
1.17 0.22  
0.80 0.17  
0.43  
0.24  
Table V. Parameters were used for INAA analysis of Selenium in biological sample by  
using 77mSe and 75Se isotopes.  
Radionuclide  
Half-life  
75Se  
77mSe  
120 d  
17.4 s  
Activation  
20 h at 3.5 x  
1012 (n/cm2/s)  
1525 s at 4.2  
x 1012  
(n/cm2/s)  
Decay time  
Counting time  
Detection limit  
20 d  
23 h  
20 s  
25 s  
1.4 ppm  
0.6 ppm  
Sample: IAEA  
436  
4.63 ± 0.48  
ppm  
4.63 ± 0.48  
ppm  
The results  
4.35 ± 1.1 ppm  
4.19 ± 0.46  
ppm  
column and channel No.13-2 were also  
determined in order to establish analytical  
procedures using the k0-NAA method. The  
system was applied to determine the  
concentration of Se in the biological sample by  
using the short-lived nuclide 77mSe. The results  
obtained through this research have opened a  
new possibility on using INAA technique for  
measurement of extremely short-lived nuclides  
at Nuclear Research Institute.  
IV. CONCLUSION  
A fast pneumatic sample transfer system  
for analyzing of extremely short-lived nuclides  
by neutron activation analysis has been  
installed and operated at Dalat nuclear research  
reactor. In this study, time parameters of the  
system were calibrated, thereby reducing  
irradiation time to seconds with precision.  
Neutron spectra parameters of the thermal  
90  
HO VAN DOANH et al.  
[4] L.S. McDowell, et al., Determination of  
ACKNOWLEDGEMENTS  
Selenium in individual food items using the  
short-lived nuclide 77mSe, Journal of  
Radioanalytical and Nuclear Chemistry, Vol.  
110, No. 2, p. 519 (1987).  
This project was carried out under the  
nuclear research and development program of  
the Ministry of Science and Technology,  
Vietnam.  
[5] A. D. Becker, Characterization and use of the  
new NIST rapid pneumatic tube irradiation  
facility, Journal of Radioanalytical Chemistry,  
Vol. 233, No. 1-2, p. 155 (1998).  
REFERENCES  
[1] S.S. Ismail, A new automated sample transfer  
system for instrumental neutron activation  
analysis, journal of Automated Methods and  
Management in Chemistry, Vol. 2010, (2010).  
[6] R. Gwozdz, F. Grass, J. Dorner, Fluorine  
analysis of standard materials by short-time  
activation analysis using 20F, Journal of  
Radioanalytical and Nuclear Chemistry, Vol.  
169, No.1, p. 57 (1993).  
[2] Yong-Sam Chung, et al., Characteristics of a  
new pneumatic transfer system for a neutron  
activation analysis at the HANARO research  
reactor, Nuclear Engineering and Technology,  
Vol. 41, No. 6, p. 813 (2009)  
[7] D. Behni, et al., Combination of Neutron  
Activation Analysis, Tracer Techniques, and  
Biochemical Methods in the Investigation of  
Selenium  
Metabolism,  
Journal  
of  
[3] U.M. El-Ghawi, et al., Determination of  
Selenium in Libyan Food Items Using  
Pseudocyclic Instrumental Neutron Activation  
Analysis, Biological Trace Element Research,  
Vol. 107, p. 61 (2004).  
Radioanalytical and Nuclear Chemistry, p.439  
(1989).  
91  